Subject:
Retinal Prosthesis
Description:
_______________________________________________________________________________________
IMPORTANT NOTE:
The purpose of this policy is to provide general information applicable to the administration of health benefits that Horizon Blue Cross Blue Shield of New Jersey and Horizon Healthcare of New Jersey, Inc. (collectively “Horizon BCBSNJ”) insures or administers. If the member’s contract benefits differ from the medical policy, the contract prevails. Although a service, supply or procedure may be medically necessary, it may be subject to limitations and/or exclusions under a member’s benefit plan. If a service, supply or procedure is not covered and the member proceeds to obtain the service, supply or procedure, the member may be responsible for the cost. Decisions regarding treatment and treatment plans are the responsibility of the physician. This policy is not intended to direct the course of clinical care a physician provides to a member, and it does not replace a physician’s independent professional clinical judgment or duty to exercise special knowledge and skill in the treatment of Horizon BCBSNJ members. Horizon BCBSNJ is not responsible for, does not provide, and does not hold itself out as a provider of medical care. The physician remains responsible for the quality and type of health care services provided to a Horizon BCBSNJ member.
Horizon BCBSNJ medical policies do not constitute medical advice, authorization, certification, approval, explanation of benefits, offer of coverage, contract or guarantee of payment.
__________________________________________________________________________________________________________________________
A retinal prosthesis replaces lost photoreceptor function by transmitting external images to an array of electrodes or via light sensors placed in the epiretinal or subretinal space. The artificial retina could restore sight to patients with blindness secondary to retinal diseases, such as retinitis pigmentosa, hereditary retinal degeneration, and some forms of age-related macular degeneration. Several models of retinal prostheses are in development in the United States, Europe, and Asia. Only the Argus II system has been cleared for use by the U.S. Food and Drug Administration.
Populations | Interventions | Comparators | Outcomes |
Individuals:
· With blindness secondary to retinal diseases | Interventions of interest are:
· Retinal prosthesis | Comparators of interest are:
· Standard treatment of retinal diseases | Relevant outcomes include:
· Functional outcomes
· Quality of life
· Treatment-related morbidity |
Background
Two approaches are being explored to develop an artificial retina that could restore sight to patients with blindness secondary to retinal diseases, such as retinitis pigmentosa, hereditary retinal degeneration, and some forms of age-related macular degeneration. The first is implantation of electrode arrays in the epiretinal or subretinal space to stimulate retinal ganglion cells. A second approach is the implantation in the subretinal space of light-sensitive multiphotodiode arrays, which stimulate the remaining photoreceptors in the inner retina. Use of a multiphotodiode array does not require external image processing. The latter approach is being evaluated for degenerative retinal diseases such as retinitis pigmentosa, in which outer retinal cells deteriorate, but inner retinal cells remain intact for years.
Research in the United States began with a first-generation, 16-electrode device (eg, the Argus 16; Second Sight Medical Products), which permitted the distinction between the presence and absence of light. Three government organizations provided support for the development of the Argus II: the Department of Energy, National Eye Institute at the National Institutes of Health, and National Science Foundation. They collaborated to provide grant funding, support for material design, and other basic research for the project.
Devices in development, none of which are approved or cleared by the U.S. Food and Drug Administration, include the following.
- The Alpha IMS was developed at the University of Tübingen, and has an electronic chip design provided by the Institute for Microelectronics, Stuttgart. The second-generation Alpha IMS device has wireless power and signal transmission and is produced by Retina Implant AG (Germany). The microchip is implanted subretinally and receives input from a multiphotodiode array with 1500 elements that moves with the eye, senses incident light, and applies a constant-voltage signal at the respective 1500 electrodes. The multiphotodiode array transforms visual scenes into corresponding spatial patterns (38´40 pixels) of light-intensity-dependent electric stimulation pulses with a maximum visual field of 15°.
- The Boston Retinal Implant Project uses an external camera mounted on a pair of glasses and a 100-electrode array. The image obtained by the external camera is translated into an electromagnetic signal transmitted from the external primary data coil mounted on a pair of glasses to the implanted secondary data coil attached to the cornea. Most of the volume of the implant lies outside the eye, with transscleral cables connected to a subretinal electrode array. The Boston Retinal Implant Project is a joint effort of the Massachusetts Institute of Technology, the Massachusetts Eye and Ear Infirmary, the Veterans Affairs Boston Healthcare System, and Cornell University.
- EPIRET3 retinal implant (Philipps-University Marburg, Germany) is a wireless system that consists of a semiconductor camera on the frame of a pair of glasses and a transmitter coil outside the eye, which sends electromagnetic signals to a receiver coil in the anterior vitreous (similar to an intraocular lens), which passes them on to a receiver microchip. A stimulator chip then generates the stimulation pulses and activates a selection of 25 electrodes placed on the epiretinal surface via a connecting micro cable.
- Intelligent Retinal Implant System (Pixium Vision, Paris, France) uses an external camera integrated with a pair of glasses and linked by wire to a pocket computer. Receiver electronics connect via a scleral tunnel to an electrode array on the surface of the retina. Pixium Vision is also developing PRIMA, which uses a subretinal implant.
- Learning Retinal Implant (Intelligent Medical Implants, Zug, Switzerland) uses an external camera on the frame of a pair of glasses and wireless data and power transfer. Receiver electronics connect via a scleral tunnel to an epiretinal implant. A retinal encoder with 100 to 1000 tunable spatiotemporal filters simulates the filtering operations performed by the ganglion cell and allows individual calibration to improve each patient’s visual perception.
- The Microelectrode-STS (suprachoroidal-transretinal stimulation) system (Osaka University, Japan) places its 9-electrode retinal prosthesis in a scleral pocket with a reference electrode in the vitreous cavity. A video camera is used to detect a visual object. Because the electrodes are at a greater distance from the retina, the resolution of the image may be lower than other devices. A proposed advantage of the STS prosthesis over epi- or subretinal prostheses is the safety of the surgical procedure, because the electrodes do not touch the retina.
Regulatory Status
In 2013, the Argus® II Retinal Prosthesis System (Second Sight Medical) was cleared for marketing by the U.S. Food and Drug Administration through a humanitarian use device exemption. This exemption is limited to devices that treat or diagnose fewer than 4000 people in the United States each year. The Argus® II system is intended for use in adults, age 25 years or older, with severe-to-profound retinitis pigmentosa who have bare light perception (can perceive light, but not the direction from which it is coming) or no light perception in both eyes, evidence of intact inner layer retina function, and a history of the ability to see forms. Patients must also be willing and able to receive the recommended postimplant clinical follow-up, device fitting, and visual rehabilitation. Food and Drug Administration product code: NBF.
Related Policies
- Keratoprosthesis (Policy #111 in the Surgery Section)
Policy:
(For Medicare Advantage, please refer to the Medicare Coverage Section below for coverage guidance.)
Retinal prostheses are considered investigational.
Medicare Coverage:
Medicare Advantage Products coverage differs from the Horizon BCBSNJ Medical Policy. There is no National Coverage Determination (NCD) for Retinal prostheses. In the absence of an NCD, coverage decisions are left to the discretion of Local Medicare Carriers. Novitas Solutions, Inc, the Local Medicare Carrier for jurisdiction JL, has determined that Retinal prostheses (0100T) will be reviewed on an individual consideration basis when submitted with documentation to support medical necessity and in accordance with Medicare Humanitarian Device Exemption (HDE) guidelines. For additional information and eligibility, refer to Local Coverage Determination (LCD): Services That Are Not Reasonable and Necessary (L35094) and Novitas Solutions, Inc, Humanitarian Device Exemption (HDE).
Available to be accessed at Novitas Solutions, Inc., Medical Policy Search page: https://www.novitas-solutions.com/webcenter/portal/MedicareJL/LcdSearch?_afrLoop=90769712476969#!%40%40%3F_afrLoop%3D90769712476969%26centerWidth%3D100%2525%26leftWidth%3D0%2525%26rightWidth%3D0%2525%26showFooter%3Dfalse%26showHeader%3Dfalse%26_adf.ctrl-state%3D63y7eftob_46
[RATIONALE: This policy was created in 2005 and has been updated regularly using searches of the MEDLINE database. The most recent literature update was performed through February 26, 2019.
Evidence reviews assess the clinical evidence to determine whether the use of a technology improves the net health outcome. Broadly defined, health outcomes are length of life, quality of life, and ability to function - including benefits and harms. Every clinical condition has specific outcomes that are important to patients and to managing the course of that condition. Validated outcome measures are necessary to ascertain whether a condition improves or worsens; and whether the magnitude of that change is clinically significant. The net health outcome is a balance of benefits and harms.
To assess whether the evidence is sufficient to draw conclusions about the net health outcome of a technology, 2 domains are examined: the relevance and the quality and credibility. To be relevant, studies must represent one or more intended clinical use of the technology in the intended population and compare an effective and appropriate alternative at a comparable intensity. For some conditions, the alternative will be supportive care or surveillance. The quality and credibility of the evidence depend on study design and conduct, minimizing bias and confounding that can generate incorrect findings. The randomized controlled trial is preferred to assess efficacy; however, in some circumstances, nonrandomized studies may be adequate. Randomized controlled trials are rarely large enough or long enough to capture less common adverse events and long-term effects. Other types of studies can be used for these purposes and to assess generalizability to broader clinical populations and settings of clinical practice. The following is a summary of the key literature to date.
RETINAL PROSTHESES
Clinical Context and Purpose
The purpose of implanting a retinal prosthesis in patients who have blindness due to retinal diseases is to provide a treatment option that is an alternative to or an improvement on existing therapies.
The question addressed in this policy is; does the use of a retinal prosthesis in patients who have blindness due to retinal diseases improve net health outcomes?
The following PICOTS were used to select literature to inform this review.
Patients
The relevant population(s) of interest are patients with blindness secondary to retinal diseases, such as retinitis pigmentosa, hereditary retinal degeneration, and some forms of age-related macular degeneration.
Interventions
The treatment being considered is the use of a retinal prosthesis. The Argus II Retinal Prosthesis System (Argus II) is the second-generation device, which has 60 electrodes. The retinal prosthesis, with the electrode array, is surgically implanted in and on the eye. The system’s external components includes of a small external video camera, held on eyeglass frames, that captures images then processed by an externally worn microcomputer. These signals are transmitted to an antenna in the prosthesis, an electronics package in the superior temporal quadrant and an electrode array implanted in the back of the eye, which in turn stimulates the optic nerve. It has been suggested that future-generation devices, containing more than 1000 electrodes, will provide more detailed vision.
Comparators
The following practices are currently being used to treat blindness due to retinal diseases; medical therapies in early stage of some conditions and adaptive interventions.
Outcomes
The general outcomes of interest are symptoms, change in disease status, functional outcomes and quality of life.
Timing
Follow-up of at least 6 months would be desirable to assess outcomes.
Setting
The prosthesis is available in tertiary treatment facilities.
Study Selection Criteria
Methodologically credible studies were selected using the following principles:
a. To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs;
b. In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies.
c. To assess longer term outcomes and adverse events, single-arm studies that capture longer periods of follow-up and/or larger populations were sought.
Studies with duplicative or overlapping populations were excluded.
A 2016 technology assessment prepared for the Agency for Healthcare Research and Quality included a systematic review of the literature on retinal prostheses.1, Reviewers included studies on the Argus II, the only retinal prosthesis cleared for marketing in the United States, as well as other retinal prostheses. Outcomes of interest were visual function, visual acuity, laboratory-based visual performance measures, day-to-day function, and quality of life. In their qualitative summary of the literature on retinal prostheses, reviewers concluded that the strength of evidence was insufficient for all outcomes.
One single-arm study with 30 patients (NCT00407602) has evaluated the Argus II retinal prosthesis; numerous articles have been published on its findings and on subgroup studies conducted on some or all of the participants. The study was prospective and multicenter, with sites in the United States and Europe. It included patients with retinitis pigmentosa (U.S.) or outer retinal degeneration (Europe) who had bare light perception or no light perception in both eyes. Articles based on this study are described next.
Humayun et al (2012) reported on interim (minimum 6-month) results on 3 types of visual acuity tasks using a computer and 2 types of real-world utility tests.2, The computer tasks included square localization (locating a high-contrast white square of light on a black background), direction of motion (indicating the direction of a high-contrast bar moving across the screen), and grating discrimination (discriminating among square-wave gratings of different special frequencies presented on a monitor). Patients performed better on all 3 computer tasks with the system on than off. In terms of the 2 real-world utility tests, with the system on, subjects had a 54% success rate in finding a door compared with a 27% success rate with the device off and had a 68% success rate in following a white line on a dark floor compared with a 23% success rate with the device off. Although all subjects were able to perceive light when the system was stimulated, the Argus II did not affect full-field light perception.
Da Cruz et al (2016) reported on 3- and 5-year results of the visual acuity tests.3, Patients performed significantly better on the 3 computer tasks with the device on than off. For the simplest task, square localization, 89% (25/28) of patients tested did better with the device on and, at year 5, 81% (17/21) of patients tested did better with the device on. For grating discrimination, the most difficult assessment, 33% (9/27) of patients tested at year 3 did better with the device on and 38% (8/21) of patients tested at year 5 did better with the device on.
Ho et al (2015) reported on safety up to 3 years.4, At 3 years postimplantation, 23 serious adverse events were reported in 11 patients; the most commonly reported were conjunctival erosion (n=4), hypotony (n=4), conjunctival dehiscence (n=3), and presumed endophthalmitis (n=3). Five-year safety was reported by da Cruz et al (2016).3, As reported by da Cruz, only 1 additional serious adverse event, a case of a rhegmatogenous retinal detachment, occurred after the 3-year follow-up (»4.5 years).3, Three devices were explanted, one each at 14 months, 3.5 years, and 4.3 years after implantation. Two patients had experienced recurrent conjunctival erosion and the third experienced chronic hypotony and ptosis.
Several publications have reported on additional functional outcomes in patients participating in the Argus II study. Patients served as their own controls; performance was compared with the device in the on vs off position. Geruschat et al (2016) reported on observer-rated assessments of visual function using the multicomponent Functional Low-Vision Observer Rated Assessment, which evaluates performance of 35 tasks.5, Tasks were grouped into 4 domains: visual orientation, mobility, daily life, and interaction with others. Twenty-six (87%) of the 30 enrolled patients were included in the analysis at a mean of 36 months (range, 18-44 months) after device implantation. All patients performed significantly better (p<0.05) in each of the 4 domains with the device on vs off, ranging from 19% to 38% improvement. Twenty-four (69%) of 35 tasks had statistically significant improvements in outcomes (ie, they were easier to perform) with the device turned on vs off.
A 2013 study reported on letter and word reading at 20 months in 21 patients participating in the Argus II study.6, Correct letter reading ranged from 51.7% to 72.3% with the device on, compared with 15.3% to 17.7% with the device off. The average time to correctly identify letters with the device on ranged from 47.7 to 68.6 seconds. Subjects who successfully completed the letter identification task proceeded to the next task. Six subjects were able consistently to read letters of reduced size. The smallest letter identified was 0.9 cm for 1 subject, but preferred letter size was as much as 22.6 cm. Four subjects were able to correctly identify 2-, 3-, and 4-letter words.
Kotecha et al (2014) reported on further testing of 6 patients from one of the Argus II study sites that had at least 3 years of follow-up; reaching and grasping outcomes were assessed.7, The test consisted of picking up a white cube from a table covered with black felt and illuminated from above, and was conducted with the electrode array on, array off, and scrambled (ie, array stimulated with a random, scattered input), in a random order. Also randomized was the location of the object, which could be placed in 1 of 4 positions. To eliminate the use of any residual vision among participants, certain patients had both eyes taped shut during the test. After 4 to 6 weeks, patients were retested to examine repeatability of performance. The percentage of successful grasps was significantly higher with the device on (69%) compared with off (0%); this finding was maintained at the second visit. With the signal scrambled, success rates were 59% at the first visit and 28% at the second visit. There were no significant differences between “on” or “scrambled” conditions for movement onset, time to object contact, or path deviation ratio, which was defined as the “deviation of the movement trajectory from a straight route between the starting and object contact wrist positions.”
Dagnelie et al (2017) evaluated performance on several functional tasks in 28 of 30 study participants who had been implanted with the device between 6 months and 3 years earlier.8, The 3 tasks were intended to have real-world application. Performance was compared with the retinal prosthesis device on and off. Task 1 was sock sorting; task 2 was sidewalk tracking; and task 3 was walking direction discrimination.
On all 3 tasks, subjects performed significantly better with the device on than off (p<0.05). (For the sock sorting task, results were presented in figures, hence precise data were not available.) With a cloth-covered table, subjects sorted approximately 70% of the socks correctly with the device on and 30% correctly with the device off. With a bare table, subjects sorted approximately 50% of socks correctly with the device and 30% with the device off. For the sidewalk task, subjects walked out of bounds a mean of 6.85 times with the device off and a mean of 4.93 times with the device on. For the walking direction discrimination task, 15 (56%) of 27 subjects performed significantly better than chance with the device on and 4 performed significantly better than chance with the device off. Although statistically significant, the clinical significance of the differences in performance on the 3 tasks is uncertain.
Summary of Evidence
For individuals who have blindness secondary to retinal diseases who receive a retinal prosthesis, the evidence includes a prospective single-arm study evaluating the device approved by the U.S. Food and Drug Administration and a systematic review of studies on various devices. Relevant outcomes are functional outcomes, quality of life, and treatment-related morbidity. A 2016 systematic review included studies on the Food and Drug Administration-approved retinal prosthesis as well as devices unavailable in the United States; the overall conclusion was that the evidence on retinal prostheses is insufficient on all outcomes of interest. One study with 30 patients has evaluated the single Food and Drug Administration-approved device (Argus II); numerous articles on this study have been published. Primary outcomes included 3 computer-based visual acuity tests. At 3- and 5-year follow-up visits, patients performed significantly better on the 3 computer tasks with the device on vs off. Performance on the most difficult task (grating discrimination) was still relatively low with the device on. Subgroup studies have tested performance on more practical tasks. These studies have tended to find significantly better performance with the device on but differences between groups may not be clinically meaningful. The same 30 patients have been evaluated multiple times and, as a result of multiple testing, their performance may differ from other individuals with the device. Additional prospective studies and additional evaluations of the ability to perform practical tasks that have a clinically meaningful impact on health outcomes are needed. The evidence is insufficient to determine the effects of the technology on health outcomes.
SUPPLEMENTAL INFORMATION
Practice Guidelines and Position Statements
No guidelines or statements were identified.
U.S. Preventive Services Task Force Recommendations
Not applicable.
Ongoing and Unpublished Clinical Trials
Some currently unpublished trials that might influence this review are listed in Table 1.
Table 1. Summary of Key Trials
| NCT No. | Trial Name | Planned Enrollment | Completion Date |
| Ongoing |  |  |  |
| NCT01024803a | Safety and Efficacy of Subretinal Implants for Partial Restoration of Vision in Blind Patients: A Prospective Multicenter Clinical Study Based on Randomized Intra-individual Implant Activation in Patients With Degenerative Retinal Diseases | 39 | Apr 2018 |
| NCT02303288a | Post-Market Study of the Argus® II Retinal Prosthesis System - France | 18 | Nov 2018 |
| NCT00407602a | Argus® II Retinal Stimulation System Feasibility Protocol | 30 | Aug 2019 |
| NCT01864486a | Restoring Vision With the Intelligent Retinal Implant System (IRIS V1) in Patients With Retinal Dystrophy (Title in France: Compensation of Vision With the Intelligent Retinal Implant System (IRIS V1) in Patients With Retinal Dystrophy) | 20 | Nov 2019 |
| NCT01860092a | New Enrollment Post-Approval Study of the Argus® II Retinal Prosthesis System | 53 | Dec 2023 |
NCT: national clinical trial.
a Denotes industry-sponsored or cosponsored trial.]
________________________________________________________________________________________
Horizon BCBSNJ Medical Policy Development Process:
This Horizon BCBSNJ Medical Policy (the “Medical Policy”) has been developed by Horizon BCBSNJ’s Medical Policy Committee (the “Committee”) consistent with generally accepted standards of medical practice, and reflects Horizon BCBSNJ’s view of the subject health care services, supplies or procedures, and in what circumstances they are deemed to be medically necessary or experimental/ investigational in nature. This Medical Policy also considers whether and to what degree the subject health care services, supplies or procedures are clinically appropriate, in terms of type, frequency, extent, site and duration and if they are considered effective for the illnesses, injuries or diseases discussed. Where relevant, this Medical Policy considers whether the subject health care services, supplies or procedures are being requested primarily for the convenience of the covered person or the health care provider. It may also consider whether the services, supplies or procedures are more costly than an alternative service or sequence of services, supplies or procedures that are at least as likely to produce equivalent therapeutic or diagnostic results as to the diagnosis or treatment of the relevant illness, injury or disease. In reaching its conclusion regarding what it considers to be the generally accepted standards of medical practice, the Committee reviews and considers the following: all credible scientific evidence published in peer-reviewed medical literature generally recognized by the relevant medical community, physician and health care provider specialty society recommendations, the views of physicians and health care providers practicing in relevant clinical areas (including, but not limited to, the prevailing opinion within the appropriate specialty) and any other relevant factor as determined by applicable State and Federal laws and regulations.
___________________________________________________________________________________________________________________________
Index:
Retinal Prosthesis
Subconjunctival Retinal Prosthesis
Second Sight Retinal Prosthesis
Argus
Argus I
Argus II
Retinal Implant
EPIRET3
Microelectrode-STS
Alpha IMS
Alpha-IMS Subretinal Implant
Turbingen Subretinal Implant
Learning Retina Implant
Suprachoroidal-Transretinal Stimulation System
Boston Retinal Implant
References:
1. Fontanarosa J, Treadwell JR, Samson DJ, et al. Retinal Prostheses in the Medicare Population (AHRQ Technology Assessment). Rockville, MD: Agency for Healthcare Research and Quality; 2016.
2.Humayun MS, Dorn JD, da Cruz L, et al. Interim results from the international trial of Second Sight's visual prosthesis. Ophthalmology. Apr 2012;119(4):779-788. PMID 22244176
3.da Cruz L, Dorn JD, Humayun MS, et al. Five-year safety and performance results from the Argus II Retinal Prosthesis System Clinical Trial. Ophthalmology. Oct 2016;123(10):2248-2254. PMID 27453256
4.Ho AC, Humayun MS, Dorn JD, et al. Long-term results from an epiretinal prosthesis to restore sight to the blind. Ophthalmology. Aug 2015;122(8):1547-1554. PMID 26162233
5.Geruschat DR, Richards TP, Arditi A, et al. An analysis of observer-rated functional vision in patients implanted with the Argus II Retinal Prosthesis System at three years. Clin Exp Optom. May 2016;99(3):227-232. PMID 26804484
6.da Cruz L, Coley BF, Dorn J, et al. The Argus II epiretinal prosthesis system allows letter and word reading and long-term function in patients with profound vision loss. Br J Ophthalmol. May 2013;97(5):632-636. PMID 23426738
7.Kotecha A, Zhong J, Stewart D, et al. The Argus II prosthesis facilitates reaching and grasping tasks: a case series. BMC Ophthalmol. May 23 2014;14:71. PMID 24885164
8.Dagnelie G, Christopher P, Arditi A, et al. Performance of real-world functional vision tasks by blind subjects improves after implantation with the Argus(R) II retinal prosthesis system. Clin Exp Ophthalmol. Mar 2017;45(2):152-159. PMID 27495262
Codes:
(The list of codes is not intended to be all-inclusive and is included below for informational purposes only. Inclusion or exclusion of a procedure, diagnosis, drug or device code(s) does not constitute or imply authorization, certification, approval, offer of coverage or guarantee of payment.)
CPT*
HCPCS
L8608
* CPT only copyright 2020 American Medical Association. All rights reserved. CPT is a registered trademark of the American Medical Association.
_________________________________________________________________________________________
Medical policies can be highly technical and are designed for use by the Horizon BCBSNJ professional staff in making coverage determinations. Members referring to this policy should discuss it with their treating physician, and should refer to their specific benefit plan for the terms, conditions, limitations and exclusions of their coverage.
The Horizon BCBSNJ Medical Policy Manual is proprietary. It is to be used only as authorized by Horizon BCBSNJ and its affiliates. The contents of this Medical Policy are not to be copied, reproduced or circulated to other parties without the express written consent of Horizon BCBSNJ. The contents of this Medical Policy may be updated or changed without notice, unless otherwise required by law and/or regulation. However, benefit determinations are made in the context of medical policies existing at the time of the decision and are not subject to later revision as the result of a change in medical policy